What is the Port Reference Impedance Conversion? Is this function available on the ENA / ENA-L?
(SUPPORTED MODELS IN BOLD: E5071C E5071B/70B E5070A/71A E5061A/62A E5061B E5072A)
The Port Reference Impedance Conversion is one of a feature of the Fixture Simulator. This function is only available on the ENA.
Using this function, S-parameter measured with system impedance (50ohm) is converted into S-parameter measured with specific arbitrary impedance.
Converting Port Impedance:
- Press the Analysis key, then click Fixture Simulator > Port Z Conversion.
- Specify the port reference impedance for the specific
- Click Port Z Conversion to change the port impedance conversion function to the ON state.
- Click Return.
- If Fixture Simulator is OFF, click the key again to enable the Fixture Simulator to the ON condition.