A noisy trace is displayed at low frequencies in the network analyzer mode of the 4395A. Is there any method for getting less trace noise?
Change the IFBW to a low BW (narrow bandwidth) setting.
|
2009-04-13
|
FAQ
|
|
|
|
|
Anomalies connecting Agilent Intuilink 4395A to the 4395A, 4396B Analyzers
|
2009-07-23
|
FAQ
|
|
|
|
|
Are there any balun transformers that can be used with the 4395A?
Yes.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can I get any instrument driver (a Plug & Play driver) for the 4395A?
No, instrument driver for the 4395A is no longer supported and is currently unavailable. Please use GPIB commands to control the 4395A.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can I use Gate Trigger mode in network or impedance analyzer mode of the 4395A?
No, the Gate Trigger mode (level / edge) can only be selected in spectrum analyzer mode of the 4395A with Option 1D6.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can I use the 1141A 200 MHz Differential Probe with the 4395A?
Yes, but you need to use the 1142A Probe Control and Power Module along with the 1141A in order to supply a DC operating power and control some input functions of the 1141A.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can I use the 16092A with the 4395A/4396B/4287A/E4991A?
Yes, you can use the 16092A with the 4395A/4396B/4287A/E4991A.
|
2009-01-14
|
FAQ
|
|
|
|
|
Can I use the 35677A (S-Parameter Test Set for the 3577A/B) for the 4395A?
No, the 4395A can't control the 35677A. The 35677A can be used only for the 3577A/B. Recommended S-parameter test set for the 4395A is 87511A/B.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can the 4395A measure integral channel power and adjacent channel power?
Yes. The 4395A programming manual provides sample programs for the power integral calculation and adjacent channel power calculation.
|
2008-06-12
|
FAQ
|
|
|
|
|
Can the 4395A put a marker on one of the traces displayed by the overlay trace function?
No, the marker can't be placed on the traces displayed by the overlay trace function of the 4395A.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can the 4395A save all of the overlay traces to one file?
No, the 4395A can't save the overlay trace data into a floppy disk or memory disk.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can the 4395A save the display screen as a color TIFF (Tagged Image File Format) file?
Yes, the 4395A can save the display screen as a color TIFF file.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can the 4395A/4396B read the cal data from the disk that comes with the cal kits such as the 85033D, 85039A, 85039B, etc?
No, the 4395A/4396B can't read the cal data from the disk furnished with these cal kits.
|
2008-11-18
|
FAQ
|
|
|
|
|
Can the spectrum analyzer mode of the 4395A display noise level below 0.0001 µV / sqrt (Hz) in noise measurement mode?
No, the 4395A can't display noise level below 0.0001 µV / sqrt (Hz) in noise measurement mode.
|
2008-11-16
|
FAQ
|
|
|
|
|
Can the time gated measurement function of the 4395A also be used in noise spectrum measurement mode?
Yes, the time gated measurement function can also be used in the noise spectrum measurement mode.
|
2008-06-12
|
FAQ
|
|
|
|
|
Do you have a GPIB command comparison chart between the 4195A and the 4395A?
Yes, please refer to 4195A-4395A GPIB Command Correspondence Table.
|
2008-11-17
|
FAQ
|
|
|
|
|
Does the 4395A have any GPIB command to extract a graphic trace data and transfer it to an external PC?
There is no command for the 4395A to save a graphic image file (TIFF format) into an external PC.
|
2008-06-12
|
FAQ
|
|
|
|
|
How can I change system impedance of the 4395A to a desired value from 50 ohms?
Yes, you can set the 4395A system impedance to your desired value by using "SET ZO" function in the calibration function menu.
|
2008-06-12
|
FAQ
|
|
|
|
|
How can I clear an error message shown on the display of the 4294A/4395A/4396B by using GPIB command?
Send the "KEY 18" command to the instrument.
|
2009-01-08
|
FAQ
|
|
|
|
|
How can I decode binary data with an external PC after I save measurement data in binary format into floppy disk on the 4395A?
No, Agilent does not provide the method of decoding the binary data saved in memory (flexible disk or memory disk) on the 4395A directly.
|
2008-06-12
|
FAQ
|
|
|
|
|
How can I derive the following equation which converts the S21 to the transmission impedance (Zt), shown in page A-14 of the 4395A Operation Manual?
The equivalent circuit model of transmission measurement is as follows.
|
2008-11-17
|
FAQ
|
|
|
|
|
How can I find accessories for an Agilent product?
For instrument-specific accessories (such as probes for an oscilloscope), first go to the product page and then see the related accessories.
For general-purpose accessories (such as rf & microwave test accessories, carts, and racks), go to the T&M Accessories, Cabinets & Cables page.
|
2008-11-21
|
FAQ
|
|
|
|
|
How can I measure the input impedance of a balanced DUT by using the 4395A?
Please refer to application note.
|
2008-11-16
|
FAQ
|
|
|
|
|
How can I measure the return loss and insertion loss of a balanced DUT by using the 4395A?
In order to measure the return loss or insertion loss of a balanced DUT, balun transformers are required.
|
2008-11-16
|
FAQ
|
|
|
|
|
How can I order the upgrade kit of the 4395A
The model numbers of available upgrade kits for the 4395A options are as follows.
|
2008-11-17
|
FAQ
|
|