Can I use regular AAA batteries in the U1241A and U1242A Handheld Digital Multimeters?
Yes, you may use regular AAA batteries for the U1241A and U1242A Handheld Digital Multimeters….

FAQ 2008-09-24

What type of internal memory is installed into the 34420A NanoVolt, Micro-Ohm Meter?
Please refer to the document linked to this page for a comprehensive listing of memory installed into the Agilent 34420A.

FAQ 2008-04-09

Can I measure temperature with the Agilent 34401A?
Most temperature sensors convert temperature to a corresponding resistance (thermistors and RTDs) or low-level dc voltag (thermocouples).

FAQ 2008-04-03

What is the Temperature Measurement Range of the Agilent 34307A Thermocouples?
Use the specifications of the Omega Engineering Inc thermocouple wire with part number TT-J-24.

FAQ 2007-01-09

What types of memory are installed into the Agilent 34401A Digital Multimeter?
Please refer to the document linked to this page for a comprehensive listing of internal memory for the 34401A digital multimeter.

FAQ 2006-11-03

What is Option 001 for the Agilent 3458A and is it a retrofittable option?
Option 001 expands reading memory from 20K to 148K (from 10,240 16 bit readings to 65,536 16 bit readings). This is a field installable option. Installation of this option does not affect the instrument calibration.

FAQ 2006-01-30

Does the Agilent 34401A use "AC" or "DC" coupling for AC voltage measurements?
The Agilent 34401A uses an AC-coupled input when measuring ACV, which means that the DC component is blocked and is not included in the measurement.

FAQ 2005-05-02

What is the sample rate of the Agilent 34401A?
When configured for the fastest reading rate, the 34401A can sample at 1000 readings/second.

FAQ 2005-02-09

How do I calculate total measurement error on the Agilent 34401A?
The accuracy specifications for the Agilent 34401A are expressed in the form: '% of reading + % of range' which match the dominate sources of error, gain and offset.

FAQ 2005-02-09

What is aperture time of the 34401A?
Aperture time is analogous to integration time or Number of Power Line Cycles (NPLCs).

FAQ 2005-02-09

Does the Agilent 34401A have a LAN IO Connection?
No. However, if you require Local Area Network (LAN) connectivity, consider the Agilent E5810A LAN-to-GPIB gateway.

FAQ 2004-09-23

Is there a way to measure low-level ac current using the Agilent 34401A?
Yes. The lowest specified measurement on the 1 A ac current range is 1% of the range, or 10 mA (see footnote 4 in the Data Sheet).

FAQ 2004-06-10

How do I view min-max values with the Agilent 34401A?
You can store the minimum and maximum readings during a series of measurements. The following discussion shows how to read the minimum, maximum, average, and reading count.

FAQ 2003-12-04

What is the part number for the User's Guide and Service Guide for the Agilent 34401A?
The part number for the English set of manuals including both the User's Guide and Service Guide is 34401-90100. The User's Guide and Service Guide are not sold separately.

FAQ 2003-11-04

How do I enable the Agilent 34401A to make dBm measurements over the bus?
First, select the dBm function using the following command: CALCulate:FUNCtion DBM

FAQ 2003-06-04

Is the current path uninterupted during measuement range changes in the Agilent 34401A?
An open circuit does not occur when range changes are made during an autorange function. When the range changes, a 0.1 Ohm resistance is switched in parallel with the current shunt resistor.

FAQ 2003-06-04

Why do I see a real value during a 4-wire ohm measurement when the test leads are open on the Agilent 34401A?
When making 2 wire ohm measurements the source and sense leads are connected together internally and can be driven into an overload situation.

FAQ 2003-06-04

Can I set the test current when making resistance measurements with the Agilent 34401A?
No, the test current is set depending on the range of the instrument. The specifications are the following:

FAQ 2003-06-04

When performing a measurement count query (SAMPLE:COUNT?) command, are the math functions reset in any way?
No. The stored values are only cleared when you turn the min-max function off, turn off the power, or perform a remote interface reset.

FAQ 2003-06-04

What are the part numbers for rack mounting the Agilent 34401A?
In order to maintain a consistent appearance across all Agilent Technologies products, the exterior color of the Agilent 34401A was changed in late 1996.

FAQ 2003-06-04

Can I measure duty cycle or pulse width with the 34401A?
No, the 34401A was not designed to measure either duty cycle or pulse width.

FAQ 2003-06-04

How do I use the calibration message function of the Agilent 34401A?
You can use the calibration message feature to record calibration information about your multimeter.

FAQ 2003-06-04

Is there a carrying case available for the Agilent 34401A multimeter?
Yes, there is both an accessory pouch and a basic instrument transit case available for the 34401A.

FAQ 2003-06-04

What is the voltage limit that is considered open for diode testing?
The 34401A diode test function was designed specifically for silicon diodes. The cutoff voltage for diode testing is 0.8 VDC.

FAQ 2003-06-04

What are the advantages of 4-wire resistance measurements with the Agilent 34401A?
Lead resistance is not included in the measurement results when using four wire ohms. Thus the accuracy is higher especially for small resistances.

FAQ 2003-06-04