Where does V3500A source the power from when connected USB port with battery installed?
USB port power

FAQ 2010-10-07

Can I use N432A thermistor power meter with S486A/G486A/J486A/H486A/X486A/M486A/P486A/K486A/R486A?
Can,but the compatibility of the N432A with those legacy sensors are not warranted.

FAQ 2010-09-22

Can the sensor cable of N1921A/N1922A wideband power sensor be extended?
The sensor cable of N1921A/N1922A cannot be extend.

FAQ 2010-09-21

Is N432A power meter drop-in replacement for legacy 432A power meter?
No, the N432A is not a drop-in replacement for legacy 432A. Refer to Migration Guide.

FAQ 2010-09-20

What is the smallest frequency step size(resolution) for V3500A handheld power meter?
1MHz step size(resolution) can be done via programming method.

FAQ 2010-09-20

What is the function of the built-in range calibrator in the N432A?
The built-in range calibrator is used to supply a DC source to the DC Bridge during calibration.

FAQ 2010-09-20

Why can’t I make a gated measurement for a pulse width wider than 45 ms?
The current version of firmware can make gated measurements with a total time of 150 ms. However the gate length can not be greater than 45 ms.

FAQ 2010-09-20

What average battery lifetime of lithium battery (1420-0394) which is located on N432A, N1911/12A, N1913A/N1914A & N8262A motherboard?
5 years

FAQ 2010-08-29

What should I do when seeing crash during the installation of Power Analyzer (software licensing version) into my non-English version of Windows…
The workaround steps to resolve this are listed below….

FAQ 2010-08-09

I get the error message “Agilent U2000 Upgrade Utility has encountered a problem and needs to close” when I run the Agilent U2000 Upgrade Utility…
This problem may be due to a Microsoft ‘.NET’ problem. We recommend that .NET be removed, and then reinstalled onto the computer.

FAQ 2010-08-09

When a B-model or H-model of the E-Series Power Sensors (with a 30 dB attenuator attached) is connected to a EPM Series, EPM-P Series or …
When a B- or H-model of the E-Series Power Sensors is connected to a EPM, EPM-P or P-Series Power Meter, the power level is 0 dBm because…..

FAQ 2010-08-09

Why am I getting this message: "Certain features in the instrument firmware are currently not supported. Please refer to Agilent Web FAQ"…
The message "Certain features in the instrument firmware are currently not supported. Please refer to Agilent Web FAQ" appears because....

FAQ 2010-08-09

What should I do when my Power Analyzer (software licensing version) still goes into trial version even though a valid license was installed…
When this situation happens, try the following steps to get this resolved:

FAQ 2010-08-09

Why do the error messages "1105: Could not connect to License Server. Please restart application" and "1003: Failed to connect …
The error messages appear when the N1918A Power Analysis Manager is launched because the Agilent License Service sometimes does not start….

FAQ 2010-08-09

When I connect the U2004A USB power meter to my device I get a low level power measurement, even when there is no RF applied …
The U2004A power sensor is DC coupled. Any low frequency low level signal on your device may cause a low level signal.

FAQ 2010-08-09

Which is power source will be utilized by V3500A power meter if external power source connected, USB cable connect to USB port & with battery insta...
External power source

FAQ 2010-07-18

How I upgrade or downgrade the V3500A handheld power meter firmware?
V3500A firmware upgrade or downgrade only can be done by Agilent only.

FAQ 2010-07-18

Can I use external power adaptor to charge my V3500A handheld power meter?
No.

FAQ 2010-07-18

What is the correct linearity setting for V8486A & W8486A sensor?
The automatic selection must be overridden: select the D type linearity for this sensor.

FAQ 2010-05-10

Can I use USB cable to charge my V3500A handheld power meter?
No

FAQ 2010-05-10

What is the method to power up V3500A handheld power meter?
You can use battery power, USB port or external power supply/adaptor to power up V3500A handheld power meter.

FAQ 2010-05-10

Can U2004A USB power sensor perform burst/time gated power measurement?
No

FAQ 2010-05-10

What is the calibration interval for 432A power meter?
1 year

FAQ 2010-05-05

Is N8262A compatible with Agilent Intuilink software?
N8262A is not compatible with Agilent Intuilink software.

FAQ 2010-03-22

Is E4416A compatible with Agilent Intuilink software?
E4416A is not compatible with Agilent Intuilink software.

FAQ 2010-03-22