Can I read a list sweep measurement data by operating the soft-panel of the E4980A on Web Browser?
Yes, you can read a list sweep measurement data using Web-server control mode.

FAQ 2010-09-19

External test, 26 TRD LOSS test fails. How do I execute?
In the 26 TRD LOSS test, use 2 m N(m)-N(m) cable for the connection between S and R on the 4291B front panel. If it fails even using the 2m cable, troubleshoot the test station.

FAQ 2010-09-19

4263B: What is the signal source output impedance?
*

FAQ 2010-09-08

When utilizing Save/Recall on the E4991A file names are limited to six characters. Is there a method to save/recall files with longer names?

FAQ 2010-07-12

Why isn't calibration option (UK6) available for the 42941A?
The 42941A does not have calibration option because the 4294A specifies total measurement accuracy including the 4294A accuracy.

FAQ 2010-07-12

How can I replace the contact tip of the 16334A?
The replacement contact tip is orderable by part number 16334-09002.

FAQ 2010-07-02

Can the E4980A perform open/short/load correction at user-specified spot frequency points while open/short correction at all frequency points is OF...
No, open/short/load correction at user-specified spot frequency points can be activated only when open/short correction at all frequency points is set to on.

FAQ 2010-07-02

Why is Rp or G measurement value unstable when pure water is measured with the 16452A?
There are two conceivable causes of unstable measurement results. One of primary causes is a change in temperature of pure water sample measured. The other is a slight pollution which resides on the electrode surfaces.

FAQ 2010-07-02

The Load correction status displayed on the 4263B LCD doesn't change to ON or OFF in response to the [:SENSe]: CORRection[:STATe] command.
Upgrade the firmware if it's older than Rev 1.05. Otherwise, take the following methods to control the correction mode.

FAQ 2010-07-02

Do the coaxial Short and 50 Ω Load terminations for the 42941A have accuracy specifications?
No, both the coaxial Short and 50 Ω Load standard terminations for the 42941A do not have accuracy specifications.

FAQ 2010-07-02

Why does the DC resistance measurement function of the 4287A require to be calibrated at the same time with Open/Short/Load calibration?
The specified accuracy for the DC resistance measurement is applied when the Open/Short/Load calibration for the DC resistance measurement is performed.

FAQ 2010-07-02

Why cannot I control TEST SET ATTEN PT1 and PT2 functions in ATTENUATOR MENU when the 87511A S-parameter test set is used with the 4395A?
The TEST SET ATTEN PT1 and TEST SET ATTEN PT2 functions cannot be controlled because the 87511A is not equipped with input attenuators.

FAQ 2010-07-02

The E4980A/4284A display much lower resistance values than DUT's DC resistance when measuring high-value resistors in Mega-ohms region.
Set the measurement function to Cp - Rp mode to measure high resistance properly.

FAQ 2010-07-02

How wide range of resistivity can the 4339B measure when used with the 16008B?
The 4339B can measure surface resistivity up to 7 x 1017 Ω and volume resistivity up to 8 x 1018 Ω-cm, but it depends on the electrodes of the 16008B used.

FAQ 2010-07-02

Does typical permittivity measurement accuracy includes all the uncertainties due to the E4991A with opt 002, 16453A and MUT insertion conditions?
Yes. All the uncertainties are included in the typical measurement accuracy.

FAQ 2010-07-02

Why does measurement data exhibit a discontinuity at 5 MHz or 15 MHz and how can the 4294A get rid of this phenomenon?
The 4294A extends its measurement frequency range up to 110 MHz by terminating each measurement terminal with 50 ohm in order to eliminate the resonance of test leads (including leads inside the 4294A).

FAQ 2010-07-02

How can I get DC bias current resolution of 20 mA in the 1 A to 2 A range when the E4980A-002 is used with 2 units of the 42841A?
The bias current resolution has been improved on the E4980A with firmware revision 2.12 and above. If your E4980A has firmware revision 2.10 or below, upgrade the firmware to revision 2.12 or above.

FAQ 2010-07-02

What is the 4294A's GPIB command to read sweep frequency data though OUTPDATA command transfers only measurement data?
"OUTPSWPRM" or "OUTPSWPRMP" are the commands to read out the sweep parameters according to the number of points.

FAQ 2010-07-02

Is there any effective way to measure the impedance of a pressure sensitive SMD component (ex. a ferrite bead) with the 16196 series or 16197A test...
In order to measure a pressure sensitive device, the pressure applied to the DUT by the test fixture needs to be carefully controlled.

FAQ 2010-07-02

Is it possible to separate the test head from the E4991A and interconnect them with cables?
Yes. It is possible with option E4991A-010 or option E4991A-007 only.

FAQ 2010-07-02

Can 4263B preserve Open/Short/Load correction data by using Save/Recall function?
No.

FAQ 2010-07-02

Can I have criteria or a list of non-applicable liquid materials for the 16452A?
Avoid measuring acidic, alkaline and ionic materials with the 16452A.

FAQ 2010-07-02

How can I order the 4263B option upgrade kit?
Refer to "4263B LCR Meter Option Upgrades" for more details of the upgrade kits.

FAQ 2010-07-02

Why is Sweep Delay Time of the 4294A prolonged when DC bias is set to ON?
The DC bias source inside the 4294A requires a certain settling time before starting a sweep measurement.

FAQ 2010-07-02

Can the 4294A measure a grounded DUT?
The 4294A can measure the grounded DUT when used with the 42941A Impedance Probe.

FAQ 2010-07-02