Contact an Expert

Technical Support

Test & Measurement

Find by Product Model Number: Examples: 34401A, E4440A

26-50 of 510

Sort:
What is required to use Remote UI for the E4991A?
You need to prepare the following items: the E4991A, a PC, the E4991A VBA software (E4991A_vba.msi) included in the furnished CD-ROM and, LAN environment. For details, please refer to chapter 10 "Using LAN" of the E4991A operation manual.

FAQ 2012-05-10

Why is a negative D value sometimes displayed when a furnished PTFE plate is measured with the 16453A?
When resistance value of Short compensation data is too large, measured D value is overcompensated for the Short resistance and results in a negative value

FAQ 2012-04-08

On the ENA 6 series, can I run a VBA program which is made on the ENA 7 series?
A VBA project for the ENA 7 series doesn't work on the ENA 6 series as is. However, the program is likely to work with exporting and importing of the modules.

FAQ 2012-03-18

How can the memory in the ENA be erased for security purposes?

FAQ 2012-03-16

Does the ENA 6 series have the swept sweep mode in the frequency sweep?
No. The E5061A/62A and E5061B have only stepped sweep mode in the frequency sweep.

FAQ 2012-03-16

On the ENA 6 series, is the User Defined BEGIN key available like the 871x?
The "Load & Run" of the E5061A/62A and E5061B are equivalent to the User Defined BEGIN key.

FAQ 2012-03-16

On the ENA 6 series, can I get TTL signal indicating Pass/Fail of the limit test like the 871x?
Yes.

FAQ 2012-03-16

Does the ENA 6 series support barcode reader?
No.

FAQ 2012-03-16

Is it possible to generate touchstone files on the ENA?

FAQ 2012-03-16

Does the ENA 6 series support mixer measurements like the 871x?
No.

FAQ 2012-03-16

Does the ENA allow user access to the Error Coefficient Arrays?

FAQ 2012-03-16

Do the ENA provide a probe power connection for active probes?

FAQ 2012-03-16

Is the display intensity adjustable on the ENA?

FAQ 2012-03-16

What types of interfaces are available on the ENA?
Please refer to the manuals.

FAQ 2012-03-16

Does the ENA support USB mouse or keyboard?
It depends on the models.

FAQ 2012-03-16

Is the ENA 6 series code compatible with the 871x series?
No, but 871x to ENA 6 series code conversion kit is available to assist the code conversion.

FAQ 2012-03-16

On the ENA I am receiving the message "Continuous switching may damage source attenuator". What corrective actions should I pursue?

FAQ 2012-03-16

What is the recommended transit case for my Agilent ENA Series Network Analyzer?
Custom fit ENA Series transit cases are not available. However customer can refer to other commercial company’s product like Pelican or ECS case.

FAQ 2012-02-15

How is trace math implemented within the Agilent Technologies Vector Network Analyzers?

FAQ 2012-01-11

What are the implications of swapping test heads between different Agilent 4287A Impedance Analyzers?
The Agilent 4287A Impedance Analyzer must be used with the furnished test head as a set because it is an indivisible portion of the calibrated 4287A.

FAQ 2011-12-19

What is an alternative solution to the 87075C / 87050E replacement?
Build your home-made switch box using the L4491A RF/MW switches. Refer to the application note for details.

FAQ 2011-12-09

Does the ENA support 87050E / 87075C series multiport test sets?
Unlike the E5061A and E5062A, the E5061B, E5071C and E5072A (ENA) does NOT support full functions of 87050E / 87075C. But it is possible to change switch state of 87050E-Hxx / 87075C-Hxx via GPIB.

FAQ 2011-12-08

ENA & PNA Series Network Analyzer Frequently-Asked Questions (FAQs)

FAQ 2011-10-07

On the ENA and ENA-L, is there a SCPI command to query the active channel?
Yes.

FAQ 2011-10-07

Does the ENA-L support a built-in web page similar to the 871x family?

FAQ 2011-10-07