Can I use any instrument driver software for the 4284A?
National Instrument’s LabVIEW driver supports the 4284A.
2009-03-11 FAQ
Can I use the 16048B with the 4285A for the cable extension?
Yes, the 16048B (Part Number 16048-60030) can be used with the 4285A. But, it is not easy to perform the cable correction using the 100-ohm resistor box (furnished with the 4285A).
2009-01-13 FAQ
Do the 4284A GPIB commands comply with SCPI?
Yes. The GPIB programming language of the 4284A is SCPI.
2009-03-11 FAQ
How can I order supplemental parts for component test fixtures/accessories?
You can order the necessary parts by the part numbers shown in the Accessories Selection Guide for Impedance Measurements.
2008-04-22 FAQ
How can I perform the calibration/compensation when I use the 16094A (part number 16094-65000) with a 7-mm connector extension cable?
It is recommended to perform the following compensation procedure...
2009-01-13 FAQ
How Can I use the Handler Interface (Opt. 201)? 2002-06-27 FAQ
How do I calculate the accuracy of 4284A when the desired stimulus is RMS current rather than RMS voltage?
The Agilent 4284A data sheet provides measurement accuracy as a function of stimulus level in RMS volts.
2007-05-09 FAQ
How do I get Instrument Drivers for Agilent Test & Measurement Products?
From the CD shipped with the instrument or from the product technical support web page for the instrument.
2009-05-01 FAQ
How would I distinguish between old EEPROM (PN 1818-5719) and new one (PN 5183-4149) when transferring the EEPROMs to new CPU board?
It can be distinguished with the manufacturer part number on the top of the EEPROMs.
2008-06-27 FAQ
Is it possible to measure the leakage inductance of a transformer via an LCR meter or impedance analyzer?
Yes, it is possible to measure the leakage inductance of a transformer via an LCR meter or impedance analyzer.
2008-01-29 FAQ
Is there a 16060A-like transformer test fixture usable with the 4284A?
No transformer test fixture for the 4284A is available.
2009-03-11 FAQ
My 16099A test fixture adapter cannot be connected to the 16191A/16192A/16193A.
An old version of the 16099A cannot be connected to the 16191A/16192A/16193A. A modification is required.
2008-05-11 FAQ
My Agilent 4284A / 4285A LCR Meter is indicating an "ADC ERR" error message. What are the corrective actions?
The error message is an indication that the LCR meter's A/D converter within the measurement circuit is malfunctioning.
2007-05-03 FAQ
On the Agilent 4284A LCR meter, how do I determine which options are installed?
2006-10-17 FAQ
Performance Test
2003-05-21 FAQ
The instrument often hangs with the error "Trap xxxx" when accessing the correction menu. What do you suggest for avoiding it?
The faulty 4284A might be recovered using a GPIB command without replacing the EEPROM or CPU.
2008-06-27 FAQ
The key cap of the power switch tends to come unfastened. What measure can I take to meet the situation?
It is recommended to replace with 0371-3953.
2008-06-30 FAQ
What are the precautions when using the Agilent 4284A Impedance Analyzer with the Agilent 42841A and 42842A/B Fixtures?
2006-08-01 FAQ
What benefits can I get by replacing the 4286A with the 4287A?
2009-03-11 FAQ
What compensation should I perform when I use the 16048B with the 4285A?
It is recommended to set the Cable Length function of the 4285A to "1 m" and perform open/short/load compensation at the DUT contact terminals of the test fixture connected to the tip of the 16048B.
2009-01-13 FAQ
What does the instrument's rear panel look like? 2002-06-27 FAQ
What does the instrument's rear panel look like? 2002-06-27 FAQ
What is a part number of the new GP-IB Assembly and ROM set?
New GP-IB assembly iis 04284-66520, and it is workable from FW Rev.03.20.
2008-06-27 FAQ
What is max. DC output voltage when two 42841A units are used as a 40A system? 2002-06-27 FAQ