85070E/85071E: What Are the Dimensions of the 7mm and 3.5mm Airlines?
|
2005-05-13
|
FAQ
|
|
|
|
|
85070E/85071E: What Are the Part Numbers for 7mm and 3.5mm Airlines for Materials Measurements?
7mm airline 85051-60010
3.5mm airline 85053-60008
|
2009-02-17
|
FAQ
|
|
|
|
|
85070E/85071E: What Are the Part Numbers for Waveguide Sample Holders for Materials Measurements?
|
2005-05-13
|
FAQ
|
|
|
|
|
Can I have criteria or a list of non-applicable liquid materials for the 16452A?
Avoid measuring acidic, alkaline and ionic materials with the 16452A. The list of non-applicable materials is not available because it is almost impossible to discriminate them from enormously large number of liquid materials existing in the world.
|
2008-05-20
|
FAQ
|
|
|
|
|
Can I replace the 16008B's conductive elastomer when it is deteriorated?
No, since the conductive elastomer cannot be removed from the main electrode assembly of the 16008B, replace whole main electrode assembly with new one.
|
2008-05-16
|
FAQ
|
|
|
|
|
Can I use somehow the 16453A for measurements up to 3 GHz?
No, the maximum usable frequency is 1 GHz because measurement uncertainty due to residual impedance of the 16453A rapidly increases at higher frequencies.
|
2008-05-20
|
FAQ
|
|
|
|
|
Can I use the 16454A with an LCR meter or Impedance Analyzer other than the E4991A-002 or 4291A/B-002?
Yes, basically, it is recommended to use an Impedance Analyzer with I-Basic programming function in order to automate the permeability calculation.
|
2008-05-20
|
FAQ
|
|
|
|
|
Can the 16453A be used with LCR meters or impedance analyzers other than the E4991A?
No, the 16453A can be used with only the E4991A and 4291A/B which have material measurement firmware (Option 002).
|
2008-04-07
|
FAQ
|
|
|
|
|
Does 85070E and 85071E software run on Windows Vista?
Windows Vista must be configured to allow the 85070E and 85071E to write files to the Program Files folder.
|
2009-02-05
|
FAQ
|
|
|
|
|
Does the 16008B comply with a certified method of national standard?
The 16008B is compatible with ASTM D257 Standard Test Methods for DC Resistance or Conductance of Insulating Materials.
|
2008-05-15
|
FAQ
|
|
|
|
|
Does the 85070E / 85071E Measurement Software supports ENA / ENA-L analyzers?
Yes.
|
2007-12-20
|
FAQ
|
|
|
|
|
Does the Agilent 85070E Dielectric Probe Kit support the PNA and PNA-L analyzers via GPIB?
|
2006-06-22
|
FAQ
|
|
|
|
|
Does the Agilent 85071E Materials Measurement Software support the PNA and PNA-L Analyzers via GPIB?
|
2006-06-22
|
FAQ
|
|
|
|
|
How can I connect a test fixture to the measurement terminals of the E4980A when the stopper (dual plastic bars) at the rear of the fixture disturbs the connection?
Remove the stopper from the test fixture by removing a screw at the center of the plastic stopper. Otherwise, remove the front rubber guide from the E4980A.
|
2009-03-11
|
FAQ
|
|
|
|
|
How can I order supplemental parts for component test fixtures/accessories?
You can order the necessary parts by the part numbers shown in the Accessories Selection Guide for Impedance Measurements.
|
2008-04-22
|
FAQ
|
|
|
|
|
How do I connect the 16048G 1m BNC connector leads to the 16452A?
You need four BNC(m) to BNC(m) adapters (Agilent part number 1250-0216) to connect the 16048G to the 16452A.
|
2009-01-12
|
FAQ
|
|
|
|
|
How do I get Instrument Drivers for Agilent Test & Measurement Products?
From the CD shipped with the instrument or from the product technical support web page for the instrument.
|
2009-05-01
|
FAQ
|
|
|
|
|
How wide range of resistivity can the 4339B measure when used with the 16008B?
The 4339B can measure surface resistivity up to 7 x 1017 Ω and volume resistivity up to 8 x 1018 Ω-cm, but it depends on the electrodes of the 16008B used.
|
2008-06-11
|
FAQ
|
|
|
|
|
Is the Agilent 85070E Material Measurement software compatible with legacy Agilent 85070A/B/C/D High Temperature Dielectric Probes?
Yes, the Agilent 85070E software is backwards compatible with all the legacy high temperature probes Agilent has sold.
|
2006-12-19
|
FAQ
|
|
|
|
|
Is the collet (center contact) of the 16454A identical to that of the 42942A (part number 1250-0907)?
Yes, When replacing the collet (center contact) of the 16454A, you can use the same part (part number 1250-0907) as that used in the 42942A.
|
2009-01-12
|
FAQ
|
|
|
|
|
There are 4 types of spacers for the 16452A. Is there any criterion for selecting the spacer?
There is no specific criterion for the selection except that the 1.3 mm spacer is required for the short compensation.
|
2009-01-12
|
FAQ
|
|
|
|
|
What are my options for running the Agilent 85070E / 85071E software to the PNA Family analyzer?
The Agilent 85070E / 85071E applications support control and communication of the PNA Family analyzers via COM or DCOM only, GPIB is not supported.
|
2006-07-18
|
FAQ
|
|
|
|
|
What are significant error sources when a thin film material is measured with the 16451B?
Small air gaps become primary error source for the permittivity measurement of thin materials.
|
2008-05-20
|
FAQ
|
|
|
|
|
What are the frequency limitations on the various probes utilized with the Agilent 85070E Dielectric Probe Kit and why?
The 85070E Dielectric Probe options 020, 030 and 050 have maximum frequeny via the E4991A Impedance Analyzer of 3 GHz.
|
2009-05-28
|
FAQ
|
|
|
|
|
What is the electrical length value defined for the 16453A?
The electrical length is 0 mm. This value is not required for measurement because calibration is performed at the electrode plane of the 16453A.
|
2008-05-20
|
FAQ
|
|