Does Agilent 89600 VSA software (89601) support the Bluetooth® Difference Error Vector Magnitude (DEVM) test?
The 89601 Option AYA does not provide DEVM results for Bluetooth basic measurement. Agilent recommends using N9081A Bluetooth measurement application.
FAQ 2012-03-27 |
|
|
How do the 89400 and 89600 families of vector signal analyzers (VSA) perform carrier-locking on a digitally-modulated signal?
First, the center frequency must be set to the carrier frequency within ±3% of the symbol rate...
FAQ 2012-02-25 |
|
|
Does the signal/spectrum analyzer support multi-user PRACH demodulation?
Yes, both the X-Series application N9082A and 89601A software can analyze multi-user PRACH. Manually add PRACH users...
FAQ 2010-05-05 |
|
|
Does the N9082A support the LTE TDD E-TM with 2-frames?
Yes. The N9082A fully supports the 2-frame (40 slots) measurements as required in the 3GPP standards for LTE TDD E-TMs.
FAQ 2009-11-13 |
|
|
Can the 89601A-BHE or N9082A be used for multi-user modulation analysis on PUSCH and PUCCH?
Yes, the 89601A-BHE and N9082A can detect the multi-user signals at one time, but for the EVM tests, they both only provide single user modulation analysis results at one-by-one.
FAQ 2009-11-13 |
|
|
What kind of Baseband filter for P-SS?
The LTE standard does not specify what transmitter filter to use. For the transmitter in ADS, we use ideal low pass filter or RRC filter; for signal studio, we use ideal low pass filter to filter baseband signal, no specific filtering on P-SS. In ...
FAQ 2009-11-13 |
|
|
Which release of the 3GPP technical specifications are supported by Agilent solutions now?
Supported versions are listed here.
FAQ 2009-11-13 |
|
|
How does the N9020A MXA with N9082A do the PvT measurement?
For additional information, click here...
FAQ 2009-11-13 |
|
|
In ACLR eNB limit, LTE 3MHz bandwidth adjacent channel carrier filter BW is the same as channel BW. Is this true?
No. The bandwidth of filter should be transmission bandwidth configuration BWconfig, not channel BW.