What are the differences between the E4980A and 4284A?
Most of the E4980A functions are compatible with the 4284A's. Since some enhancements and updates of functionality are incorporated in the E4980A, there are primary differences summarized in the table.

FAQ 2014-04-07

What is the P/N of cable between E4980A and 42841A?
The current cable is p/n E4980-61640.

FAQ 2013-12-19

Where can I find manuals for discontinued and obsolete Hewlett-Packard and Agilent test equipment?
Most Agilent products have PDF manuals that can be found on product pages. Other vendors also supply old manuals.

FAQ 2013-11-22

Should I update I/O library version? (Automatic Update)
Agilent I/O Library is preinstalled in the E4982A.

FAQ 2012-10-25

PDF PDF
On the Agilent E4980A, what is the "DC Source Option 001 (±10 VDC Max)" and how is it utilized?
A summary description of the enhancements provided by E4980A option 001 - 'Power and DC Bias Enhancement Option'

FAQ 2012-10-15

How should I set measurement points in BIN sorting of E4982A?
Setting of measurement points in BIN sorting of E4982A is independent of the selection of measurement points in the measurement display (single point measurement or list measurement).

FAQ 2012-08-01

How should I set measurement points in BIN sorting of 4287A?
Setting of measurement points in BIN sorting of 4287A is independent of the selection of measurement points in the measurement display (single point measurement or list measurement).

FAQ 2012-08-01

What are the implications of swapping test heads between different Agilent 4287A Impedance Analyzers?
The Agilent 4287A Impedance Analyzer must be used with the furnished test head as a set because it is an indivisible portion of the calibrated 4287A.

FAQ 2011-12-19

Why is the "E-230 data corrupt or stale" error message occasionally displayed on the E4980A while saving a measurement data in the USB memory stick...
It is caused when the procedure for saving a data is not performed correctly or when the USB memory stick is not compatible with the E4980A.

FAQ 2011-07-24

What is a recommended USB memory device for the E4980A and the E4981A?
4 GB USB Memory (Agilent PN 1819-0637) is recommended.

FAQ 2011-07-11

How can the E4980A or the 4263B measure internal impedance of batteries?
Insert DC blocking capacitors between the measurement terminals and a battery DUT. The 16065C can substitute for the DC blocking capacitors.

FAQ 2011-05-12

Can the 16338A be used with LCR meters such as the 4263B and the E4980A?
No, the 16338A Test Lead Set doesn’t have compatibility with LCR meters. The 16338A can be used with the 4338A/B only.

FAQ 2011-05-12

How can I know the stray capacitance and residual inductance values from OPEN A/B and SHORT A/B measurement data displayed on the E4980A LCD?
The E4980A displays the open admittance and the short impedance obtained for Open/Short compensation as OPEN A, OPEN B, SHORT A and SHORT B values on the CORRECTION Setup page.

FAQ 2011-05-12

Why does the 42841A shut down with an error message display on the E4980A/ 4284A when DC bias current exceeds a certain level?
The 42841A cannot output DC bias current in excess of a level determined from the DC resistance of DUT and the maximum limit of DC output voltage.

FAQ 2011-04-21

What should I do when the "OVLD" message is displayed on the 4263B?
Check whether the selected measurement range is appropriate or not.

FAQ 2011-04-21

What are primary differences between the 4263A and 4263B?

FAQ 2011-01-06

Can I read a list sweep measurement data by operating the soft-panel of the E4980A on Web Browser?
Yes, you can read a list sweep measurement data using Web-server control mode.

FAQ 2010-09-19

4263B: What is the signal source output impedance?
*

FAQ 2010-09-08

How can I connect a test fixture to the measurement terminals of the E4980A when the stopper at the rear of the fixture disturbs the connection?
Remove the stopper from the test fixture by removing a screw at the center of the plastic stopper. Otherwise, remove the front rubber guide from the E4980A.

FAQ 2010-07-02

How can I get DC bias current resolution of 20 mA in the 1 A to 2 A range when the E4980A-002 is used with 2 units of the 42841A?
The bias current resolution has been improved on the E4980A with firmware revision 2.12 and above. If your E4980A has firmware revision 2.10 or below, upgrade the firmware to revision 2.12 or above.

FAQ 2010-07-02

Can the E4980A perform open/short/load correction at user-specified spot frequency points while open/short correction at all frequency points is OF...
No, open/short/load correction at user-specified spot frequency points can be activated only when open/short correction at all frequency points is set to on.

FAQ 2010-07-02

The 16085B 4-Terminal Pair to 7 mm (APC-7) adapter: Is it advisable to use fixture compensation standards or coaxial calibration standards?

FAQ 2010-07-02

Can I control the E4980A functions by operation on Web Browser?
Yes, you can control the E4980A over Web Browser by connecting it to your PC by LAN.

FAQ 2010-07-02

What is input impedance of the 4339B?
Nominal input resistance is 1000 Ω .

FAQ 2010-07-02

Can 4263B preserve Open/Short/Load correction data by using Save/Recall function?
No.

FAQ 2010-07-02