For the PSA Series Spectrum Analyzers, is the new two year calibration cycle applicable to all instruments?
Based on evaluation of serviced PSAs, Agilent determined that the PSA is stable enough to support a 2 year calibration cycle for all PSA instruments.

FAQ 2012-05-30

Can the E4406A or PSA Series cdmaOne personality (Option BAC) make power timing and phase measurements?
Yes. In the cdmaOne personality you will first need to change the Measure Method to Timing and Phase under the Measure Setup key and then change the view to Power Timing and Phase under the View/Trace key.

FAQ 2012-05-29

Are there any known issues when downgrading the VSA firmware?
When downgrading firmware on the E4406A there is an issue with the firmware upgrade program. You must first completely uninstall any previous installation of the upgrade program before installing an older version.

FAQ 2012-05-29

How do I remotely read demodulated symbols from the 89601A VSA software using Microsoft® Excel?
Click here for information on reading demod symbols from the 89601A software using Excel.

FAQ 2012-05-29

Why is the AC to DC power supply for my E7495A/B Base Station Test Set not working?
Click here for information about the AC to DC power supply for E7495A/B.

FAQ 2012-05-29

How do I remotely make an Adjacent Channel Power (ACP) measurement with the 89601A VSA software using Microsoft Excel?
Click here for information on controlling the 89601A software using Excel to make an ACP measurement.

FAQ 2012-05-29

How do I remotely program my PSA Series Spectrum Analyzer or E4406A VSA Series Analyzer to transfer Raw I/Q trace data using the IVI-COM driver in...
Click here for information on transfering Raw I/Q trace data using IVI-COM and .NET.

FAQ 2012-05-29

How do I remotely read status bits and set interrupts for the 89601A VSA software using Microsoft® Excel?
Click here for information on reading status bits and setting interupts.

FAQ 2012-05-29

Does the E7495A/B Base Station Test Set allow simultaneous operation of the CW and complex source in Spectrum Analyzer mode, Channel Analyzer mode,...
Yes. This support was added for A.02.10 firmware in January 2004.

FAQ 2012-05-29

What is the CDMA Codogram display on the E7495A/B Wireless Base Station Test Set?
Codogram is a three dimensional display of Walsh code versus time versus code power. Click here for more details.

FAQ 2012-05-29

How do I remotely read the summary table for a demodulated Wireless LAN signal from the 89601A VSA software using Microsoft® Excel?
Click here for information on controlling the 89601A software using Excel to read the summary table.

FAQ 2012-05-29

Does the E7495A/B Wireless Base Station Test Set support frequency-panning in Spectrum Analyzer mode?
Yes, this feature was added in firmware version A.02.10. Click here for information on frequency-panning.

FAQ 2012-05-29

How do I remotely program my PSA Series Spectrum Analyzer or E4406A VSA Series Analyzer to read ASCII trace points without using an instrument driv...
Click here for an example demonstrating how to over LAN read ASCII trace points of an entire EDGE waveform in the GSM w/ EDGE personality Power Vs. Time measurement.

FAQ 2012-05-29

How do I upgrade the E4406A VSA Series Firmware and Measurement Personalities?
The E4406A firmware and measurement personalites can be upgraded via the LAN or crossover cable. (Click on this FAQ for more information)

FAQ 2012-05-28

Where can I find technical overviews and self-guided demonstrations of the E4406A and PSA Series measurement personalities?
Technical overviews are available for the following personalities: GSM/EDGE, cdma2000, 1xEV-DO, W-CDMA, cdmaOne, NADC/PDC, Phase Noise (PSA/ESA), and Noise Figure (PSA).

FAQ 2012-05-28

Where can I find detailed specifications for the E4406A, PSA Series, ESA Series Spectrum Analyzers?
We publish both an abbreviated Data Sheet and complete Specifications Guide for these analyzers. You can find the complete specifications guide by looking under Manuals, Guides, and Notifications at the bottom of each product web page.

FAQ 2012-05-28

How do I remotely transfer screen images from my E4406A VSA Series Analyzer using Visual Basic 6?
Click here for information on transfering screen images.

FAQ 2012-05-28

What are the measurement intervals for the W-CDMA, cdma2000, and 1xEV-DO QPSK EVM measurements for E4406A and the PSA Series? Is the 12.2k Reverse ...
Click on this FAQ to get detailed information on the measurement interval ranges.

FAQ 2012-05-28

For my ESA/PSA/EMC Series Spectrum Analyzer or E4406A VSA Series Analyzer how do I serial poll to determine when an auto-alignment is complete usin...
Click here for information on performing an alignment and using serial polling to determine when it is complete.

FAQ 2012-05-28

What file formats does the 89601A VSA software support for signal waveforms?
Recorded data from the following files can be interpreted by the 89600: ASCII (.txt and .csv), Binary (89600 SDF .sdf, .dat), Binary (E3238 .cap), Binary (N5510A .bin), MATLAB 7 (or earlier, .mat)

FAQ 2012-05-28

How do I optimize my cdmaOne ACPR measurement for the third offset (2.75 MHz offset, 1 MHz bandwidth) in the E4406A and PSA Series cdmaOne personal...
By changing the default number of FFTs segments to 5 and upgrading to the 14 bit digital IF (E4406A only) you can substantially improve your results.

FAQ 2012-05-28

Does the E4406A or PSA Series support GPRS measurements?
Both the E4406A and PSA Series support GPRS measurements. For the E4406A, option BAH (GSM) or option 202 (EDGE with GSM) are required. For the PSA, option 202 (GSM with EDGE) and option B7J (Digital Demodulation hardware) are required.

FAQ 2012-05-28

How do I upgrade the E4406A VSA Series 321.4 MHz IF Output Option 300?
This option can be upgraded at your local Agilent Service Center. (Click on this FAQ for more information)

FAQ 2012-05-28

How can I upgrade my PSA Series Spectrum Analyzer to add Gated Sweep (Time gating)?
A hardware retrofit will be required before gated sweep can be used in PSA's purchased before September, 2003 (Serial Prefix less than US4336 or MY4336 ). Firmware Release A.04.12 or greater is also needed.

FAQ 2012-05-28

For the PSA Series and E4406A VSA Series 1xEV-DO measurement personality, why did the Modulation Accuracy (Composite Rho) measurement change names ...
In the cdma2000 standard, Rho is a common parameter name used to express modulation quality. However, 3GPP2 uses the term Waveform Quality in the 1xEV-DO conformance tests, and the 1xEV-DO measurement personality was updated to use the same termi...

FAQ 2012-05-28