Will there be specification, performance characteristic or usability differences when using the 11970 Series mixers on the PXA?
Yes, there will be several differences when using 11970 Series mixers with the
FAQ 2013-04-26 |
|
|
Do I need to perform a periodic front panel instrument alignment on my N9039A RF Preselector?
No, but there is a system alignment that needs to be periodically run. See FAQ content for details...
FAQ 2012-12-11 |
|
|
66xxA, E363xA and E364xA: Can Agilent Power Supplies Sink Current?
Yes! Sinking, or down-programming, is the ability to pull current into the positive terminal of a power supply.
FAQ 2012-12-07 |
|
|
4294A의 측정 신호의 출력 임피던스는 무엇 입니까?
source의 출력 임피던스는
FAQ 2012-10-22 |
|
|
What is the typical power consumption of the 86100C and 86100D Digital Communications Analyzer?
Typical measurements of 86100C and 86100D power consumption
FAQ 2012-08-06 |
|
|
What is the measurement uncertainty for the 8902A Option 50?
The only uncertainty in the 8902A Option 050 test system should be the accuracy of the attenuation source.
FAQ 2012-05-30 |
|
|
Where can I find detailed specifications for the E4406A, PSA Series, ESA Series Spectrum Analyzers?
We publish both an abbreviated Data Sheet and complete Specifications Guide for these analyzers. You can find the complete specifications guide by looking under Manuals, Guides, and Notifications at the bottom of each product web page.
FAQ 2012-05-28 |
|
|
For the PSA Series and E4406A VSA Series 1xEV-DO measurement personality, why did the Modulation Accuracy (Composite Rho) measurement change names ...
In the cdma2000 standard, Rho is a common parameter name used to express modulation quality. However, 3GPP2 uses the term Waveform Quality in the 1xEV-DO conformance tests, and the 1xEV-DO measurement personality was updated to use the same termi...
FAQ 2012-05-28 |
|
|
What features and accessories are included in the standard E7495A Wireless Base Station Test Set?
The standard instrument includes the Antenna / Cable and Spectrum Analyzer modes and an internal GPS receiver and accessories and more. Click here for more details.
FAQ 2012-05-28 |
|
|
How often do I need to run the RF Preselector alignments in my MXE?
Every 30 days or when the temperature changes by 3 degrees...
FAQ 2012-05-09 |
|
|
What are the dimensions, instrument weight, and shipping weights of the ESA family of Spectrum Analyzers?
The instrument weight and shipping weight vary for each product.
FAQ 2012-03-16 |
|
|
What field of view (FOV) magnifications are available?
The x6000 does not use the field of view (FOV) concept, it uses an autofocus technique that sizes the inspection image to the joint type being inspected. Several slices are evaluated for the sharpest region to determine the slice location.
FAQ 2012-03-12 |
|
|
What is the maximum input power that can safely be applied to input of the E7495A/B Wireless Base Station Test Set?
The maximum input level to the RF Input is +20 dBm. The maximum RF Input when using the supplied external 40 dB attenuator is +50 dBm. The E7495A/B additionally has a forward power protection switch at the RF Input connection to help prevent dam...
FAQ 2012-02-28 |
|
|
How do I make measurements on low level over-the-air CDMA signals using the E7495A/B Wireless Base Station Test Set?
E7495A/B Options 810, 811, and 812 include a pre-selector filter that attenuates out-of-band signals. This reduces the overall signal input into the E7495A/B and allows the built in auto-ranging circuit to increase the IF gain.
FAQ 2012-02-28 |
|
|
What is the power output range for the E7495A/B Wireless Base Station Test Set Continuous Wave (CW) Signal Generator, Option 500?
The output power range is -25 to -80 dBm with +/- 1 dB level accuracy. An additional 40 dB attenuation can be added using the external 40 dB attenuator that comes standard with the E7495A/B.
FAQ 2012-02-27 |
|
|
What are the Noise Figure specifications above 3 GHz for the ESA Series Spectrum Analyzer?
The specifications for instrument noise figure uncertainty is nominally the same as from 10 MHz to 26.5 GHz, especially for high gain devices under test.
FAQ 2012-02-25 |
|
|
핸드형 스펙트럼 분석기의 실제 크기와 무게는 어떻게 됩니까?
크기(W x H x D)는 318 x 207 x 69mm(12.5 x 8.15 x 2.7인치)입니다.
FAQ 2011-11-29 |
|
|
ENA / ENA-L에서 "업데이트 펌웨어" 키가 서비스 메뉴 하단에 표시되지 않습니다. 무엇이 문제입니까?
FAQ 2011-10-07 |
|
|
Is it possible to generate negative voltage using the N6700B Modular Power System?
FAQ 2011-05-31 |
|
|
Under the unit function’s ACV and DCV, the display rate of U1271/2A indicates 7 times per second, can GUI DATA LOGGER capture all 7 readings?
No, the GUI DATA LOGGER is running on different acquisition method than the unit’s display rate, therefore the fastest update rate is 1 sample per second.
FAQ 2011-05-05 |
|
|
Why are my N4010A DEVM test results different with firmware revision A.05.03.15?
Revision A.05.03.15 was a short-lived release that used a modified measurement algorithm. Please upgrade to revision A.05.03.16, which uses the correct measurement algorithm.
FAQ 2011-02-18 |
|
|
ENA Option TDR 시간 도메인 측정의 정확도는 얼마입니까?
TDR Measurement Uncertainty(커버리지 계수 = 2) = +-0.007(ECal 이후)[*1]
TDT Measurement Uncertainty(커버리지 계수 = 2) = ±0.025*X, 여기서 X는 측정된 값을 나타냄(ECal 이후)[*1]
FAQ 2011-01-04 |
|
|
Can the ESG acheive Amplitude Modulation (AM) depths of 100%?
Yes, please see AM specifications for the ESG signal generators.
FAQ 2010-12-09 |
|
|
IO Libraries Suite: What is new with Agilent IO Libraries Suite 14.0?
Agilent IO Libraries Suite provides a new and improved set of tools for establishing, verifying, and troubleshooting instrument-to-PC connections.
FAQ 2010-10-30 |
|
|
Why does my E6651A fail Source Modulated Power Level Accuracy in TME?
This may be due to a compatibility issue between the E6651A firmware version and TME.
FAQ 2010-10-14 |
|