Why does my N4010A sometimes fail EDR Relative Power on a good DUT?
This may happen if the firmware is A.05.03.10 (or newer).

FAQ 2010-02-02

Is there any special setting that I need to do in Agilent VEE before using the U2300A or U2500A DAQ with it?
Yes. In fact it is the most important thing you need to set before using the DAQ with Agilent VEE. As most instruments uses MSB byte ordering, DAQ...

FAQ 2009-12-15

Check disk often runs when the E5070B/71B or E5061A/62A power up. Why?
Shutdown process might be failed. Uninstall the unsupported software or use special shutdown process when you want to shutdown the unit.

FAQ 2009-11-17

What is the main difference between the digital output (DO) of the U2100A, U2600A, U2300A and U2500A?
The DO port of U2100A and U2600A requires external power supply to be pumped into the port...

FAQ 2009-10-20

Why do I receive an error message below ?
Generally, if you are sampling at 60kSa/s and above, you should use high sampling mode to acquire data. This is just a rough guideline. If you encountered this message even when....

FAQ 2009-08-06

N4903A: Our N4903A lost all of its licensed options. Can they be recovered?
The licenses are tied to the Host ID. Since the unit has two ethernet cards, it may be that the wrong Host ID is being used to check licensing.

FAQ 2009-08-06

How do I configure the E5813A Event Logging features in the E5813A LAN/USB Gateway Configuration Utility?
The E5813A Event Logging features are for internal diagnostic/troubleshooting of the device......

FAQ 2009-08-06

Windows runs successfully but the ENA-L firmware doesn't start. How can it be solved?
Perform system recovery and then install the firmware 3.01 or later.

FAQ 2009-08-06

Why does handheld scope’s DMM measurement frequently stall after one hour of continuous operation?
This condition indicates that the internal memory battery level has dropped to 1.4V or lower.

FAQ 2009-07-16

Why can’t my PC connect to my InfiniiVision oscilloscope via its LAN/USB/GPIB interface?
There are many possible reasons for this, but a very common problem is that the interface has not been enabled on the scope.

FAQ 2009-05-25

How do I resolve Error 631/Error 635 when it appears on the display of the 34405A Digital Multimeter during self-test?
You can resolve Error 631/Error 635 on the 34405A Digital Multimeter by performing a firmware upgrade….

FAQ 2009-03-25

Where can I find manuals for Agilent EEsof EDA products?
Complete product documentation for all of our current products can be found at www.agilent.com/find/eesof-docs

FAQ 2009-02-03

Why doesn't Windows® detect new hardware when I plug the E5813A USB Hub into the computer?
The new hardware is not detected when you plug the E5813A USB Hub into the computer because the PC driver may not be installed properly....

FAQ 2009-01-12

What can I do if my U8101A Display Tester is not able to power up?
If the power button LED light on the front panel of the U8101A Display Tester does not turn green after the power button is pressed….

FAQ 2008-11-30

Cal kit data error could occur when language setting is changed from English on the ENA.
Please change the language setting back to the default (English).

FAQ 2008-10-08

What will happen to the E5810A LAN/GPIB Gateway (with firmware revision A.01.05) when it is powered ON for 5 minutes without a LAN connection?
When the E5810A (with firmware revision A.01.05) is powered ON for 5 minutes without a LAN connection, a static IP will be assigned….

FAQ 2008-09-25

Can I change the regional setting of Windows system running in the E5052B?
No, the default setting is "English (United States)" and the E5052B can not support the other setting.

FAQ 2008-07-18

How to recover the operating system.
For details, refer to "System Installation for A27 Mass Storage", in the service manual.

FAQ 2008-06-27

My 16099A test fixture adapter cannot be connected to the 16191A/16192A/16193A.
An old version of the 16099A cannot be connected to the 16191A/16192A/16193A. A modification is required.

FAQ 2008-05-11

Why are the networks disconnected on certain actions regarding the MCF file?
To avoid inconsistent software behavior, the networks are disconnected automatically before the MCF is reloaded or a reference MCF is saved...

FAQ 2008-01-29

Why do I get a message that a help file or DLL was not found?
This message usually only occurs if you have chosen the “compact installation” setup type or “custom installation”...

FAQ 2008-01-29

FAQs: Service Support

FAQ 2008-01-18

How do I open older Agilent Logic Analyzer application files in newer versions?

FAQ 2007-06-19

My Agilent 4284A / 4285A LCR Meter is indicating an "ADC ERR" error message. What are the corrective actions?
The error message is an indication that the LCR meter's A/D converter within the measurement circuit is malfunctioning.

FAQ 2007-05-03

On my Agilent 8753ES Analyzer, I am receiving the caution: “Caution: No if found: Check R input level”. What are the corrective actions?

FAQ 2007-02-26