Is Option A6J or 1A7 calibration available on the X-Series, PSA, ESA analyzers and X-Series, ESG-C, PSG signal generators directly from the factory?
Yes, there are additional details for certain PSA configurations that can be found here.
FAQ 2013-05-17 |
|
|
How long is the factory channel correction calibration good?
The factory channel correction lasts the life of the source.
FAQ 2012-12-26 |
|
|
Do temperature changes affect the factory channel correction calibration?
temperature changes, channel correction calibration, factory correction calibration,x-series signal generators
FAQ 2012-12-26 |
|
|
Can running user correction calibration over a narrower range improve calibration?
No, the number of points is always fixed at a 5 MHz interval
FAQ 2012-12-26 |
|
|
For User channel correction calibration, when frequency is tuned out of “user” correction frequency range, is factory correction applied?
Yes, we apply smoothing between cal arrays to ensure no discontinuity.
FAQ 2012-12-26 |
|
|
How does DC calibration affect channel corrections?
If a DC calibration is run with channel corrections applied, this may or may not degrade performance slightly.
FAQ 2012-12-26 |
|
|
For User channel correction calibration, is it better to do a spot user cal or over an overlapping range?
Overlapping is good idea, but not required.
FAQ 2012-12-26 |
|
|
For User channel correction, how many unique calibration files are supported?
We can support thousands on internal solid state drive.
FAQ 2012-12-26 |
|
|
When you perform the User channel correction routine, do you need to compensate for the cable between spectrum analyzer or power meter and source?
No, the cable error has been calibrated.
FAQ 2012-12-26 |
|
|
How can I read the calibration date for an E363xA or an E364xA power supply?
The calibration message containing the factory calibration date can be retrieved by using the "CALibration:STRing" command. After the calibration is completed, the date can be changed with the " CALibration:STRing " command.
FAQ 2010-09-08 |
|
|
How long is the calibration time interval?
Agilent recommends 1 year as the calibration interval for all E3600 series power supplies.
FAQ 2008-09-23 |
|
|
66xxA: Calibration Certificates
Criteria have been established to determine which products get or do not get a calibration from the factory. A product with a product number does not automatically qualify for a calibration certificate.
FAQ 2008-09-22 |
|
|
66xxA: Re-Calibration Date
There are no components used in Power Products which have a specific shelf life. Therefore, the calibration period of this instrument begins on the date that the product is first put into service by the customer. The recommended calibration interv...
FAQ 2008-09-22 |
|
|
For the ESG and ESG-D Series, does the IQ calibration routine include inputs and outputs, and what are the limits/specifications?
FAQ 2005-07-22 |
|
|
What is the calibration interval for ESG series RF Signal Generator?
FAQ 2005-05-23 |
|