Is it possible on the Agilent ENA series Network Analyzers to over-ride the output source blanking at every sweep end?
Yes, the output source blanking on sweep retrace can be prevented by turning off the Initial Source Port Control function.
FAQ
2012-10-17
|
|
|
What is the difference between an "ECal Confidence Check" and a Network Analyzer "System Verification" test?
The ECal Confidence Check is a method to check the accuracy of a calibration performed with either a mechanical cal kit or an ECal module. System verification testing provides network analyzer users a method to verify that the network analyzer is ...
FAQ
2012-10-17
|
|
|
On the Agilent E4980A, what is the "DC Source Option 001 (±10 VDC Max)" and how is it utilized?
A summary description of the enhancements provided by E4980A option 001 - 'Power and DC Bias Enhancement Option'
FAQ
2012-10-15
|
|
|
Does the ENA 6 series have the swept sweep mode in the frequency sweep?
No. The E5061A/62A and E5061B have only stepped sweep mode in the frequency sweep.
FAQ
2012-03-16
|
|
|
How is trace math implemented within the Agilent Technologies Vector Network Analyzers?
FAQ
2012-01-11
|
|
|
What are the implications of swapping test heads between different Agilent 4287A Impedance Analyzers?
The Agilent 4287A Impedance Analyzer must be used with the furnished test head as a set because it is an indivisible portion of the calibrated 4287A.
FAQ
2011-12-19
|
|
|
Using the VBA program, can the ENA / ENA-L control the handler interface to control other peripherals?
Yes.
FAQ
2011-10-07
|
|
|
Does the touchstone file generated by the ENA / ENA-L include the electrical delay?
The touchstone files do not include electrical delay. Please use port extension.
FAQ
2011-10-07
|
|
|
The "set Zo" function does not appear to affect the displayed S-parameters of my network analyzer. How does this function work?
FAQ
2011-10-07
|
|
|
On the ENA, how can I make CW sweep (Fixed frequency) without turning off the ENA's output signal each time a sweep finishes?
Turn off the "Init Src Ctrl" feature, then hold trigger.
FAQ
2011-10-07
|
|
|
For Cable Testing should I use a TDR, a network analyzer, or both?
The cable specifications will determine the test equipment requirements...
FAQ
2011-10-07
|
|
|
On the ENA / ENA-L, what occurs if I select "order-based display" in sweep mode?
Measurement points of different segments are plotted in constant intervals.
FAQ
2011-10-07
|
|
|
What are the equations that relate Real and Imaginary data pairs to Log Magnitude, Polar, Phase, and Smith Chart formats?
FAQ
2011-10-07
|
|
|
Is it possible to change the font size of softkey menu on ENA?
Yes. The font size of softkey menu can be changed with following procedure...
FAQ
2011-10-07
|
|
|
What is the difference between Smoothing and Averaging on an Agilent network analyzer?
FAQ
2011-10-07
|
|
|
When I save a Touchstone file with the ENA / ENA-L, which calculation will be reflected in the Touchstone file?
All calculations are reflected in the Touchstone file except for the mixed-mode conversion.
FAQ
2011-10-07
|
|
|
What is the difference between "sweep speed", "cycle time" and "time per data point" as related to a vector network analyzer?
FAQ
2011-10-07
|
|
|
How can I calculate the equivalent phase delay, as a function of frequency, for a given electrical delay as set on Vector Network Analyzer?
Equivalent phase delay, based upon electrical delay is computed as: θ˚ = Ftest * Delay * 360
FAQ
2011-10-07
|
|
|
How can the S21 to transmission impedance (Zt) conversion equation on the 4395A Operation Manual be derived?
The equivalent circuit model of transmission measurement is as follows.
FAQ
2011-06-03
|
|
|
How can I know the stray capacitance and residual inductance values from OPEN A/B and SHORT A/B measurement data displayed on the E4980A LCD?
The E4980A displays the open admittance and the short impedance obtained for Open/Short compensation as OPEN A, OPEN B, SHORT A and SHORT B values on the CORRECTION Setup page.
FAQ
2011-05-12
|
|
|
For the 4395A /4396B combination analyzer in spectrum analyzer mode with the noise format. Are the indicated units in error for this mode?
FAQ
2011-05-06
|
|
|
The E4980A/4284A display much lower resistance values than DUT's DC resistance when measuring high-value resistors in Mega-ohms region.
Set the measurement function to Cp - Rp mode to measure high resistance properly.
FAQ
2010-07-02
|
|
|
Why does the DC resistance measurement function of the 4287A require to be calibrated at the same time with Open/Short/Load calibration?
The specified accuracy for the DC resistance measurement is applied when the Open/Short/Load calibration for the DC resistance measurement is performed.
FAQ
2010-07-02
|
|
|
Does typical permittivity measurement accuracy includes all the uncertainties due to the E4991A with opt 002, 16453A and MUT insertion conditions?
Yes. All the uncertainties are included in the typical measurement accuracy.
FAQ
2010-07-02
|
|
|
What are primary differences between the E4991A and 4291B?
The major difference is frequency range. E4991A: 3 GHz, 4291B: 1.8 GHz.
FAQ
2010-07-02
|
|