技术支持
1-25 / 101
|
How often does the system user alignment for the PSA EMI Measurement Receiver need to be run?
30 days or 3° C temperature change. See contents for details...
FAQ 2012-12-11 |
|
|
What is preventing the 89600A software running on a PC from connecting to an X-Series analyzer?
These requirements MUST be observed when using the 89600A VSA software with an X-Series Signal Analyzer.
FAQ 2012-06-12 |
|
|
为什么我在信号分析仪模式中进行零扫宽测量时会得到“Meas UNCAL”信息?
在进行零扫宽测量时出现“Meas UNCAL”信息,表明在当前 RBW 和扫描时间组合设置下,捕获存储器中存储的数据已经超出了其存储容量。
FAQ 2012-06-01 |
|
|
为什么 FieldFox 有时候会恢复到开机时的零序列号?
固化软件需要升级到 A.03.02 或更高版本。
FAQ 2012-06-01 |
|
|
当进行 TME 软件调整时,如何才能避免“缓冲区溢出”?
在 FieldFox 分析仪和 PC 之间使用交叉电缆直接进行连接。
FAQ 2012-06-01 |
|
|
为什么我在尝试升级固化软件时会看到“Install Error”信息?
可能是因为文件名冲突。
FAQ 2012-06-01 |
|
|
为什么在我的 N9923A 上会出现“VCO 解锁错误”?
解决办法是将固化软件升级到 A.05.04 或更新版本。
FAQ 2012-06-01 |
|
|
Why do I get a Meas Uncal message when I select either the Quasi-Peak or EMI Average detector for taking a full band sweep with the MXE EMI Receiver?
Sweeptime has reached its maximum. See complete FAQ for further details...
FAQ 2012-05-10 |
|
|
Why do I get the "Align 20 Hz to 3.6 GHz Required" message every time I turn my MXE on?
You don't. Wait for it to warm back up...
FAQ 2012-05-09 |
|
|
Why do I continue to get the "Align 20 Hz to 3.6 GHz Required" message even though I ran the Align Now, All routine on my MXE?
You also need to run the System, Alignments, More, RF Preselector, Align Now alignments...
FAQ 2012-05-09 |
|
|
How long does the system user alignment of my PSA EMI Measurement Receiver take to run?
A full system alignment takes approx. 30 minutes. See FAQ content for further details...
FAQ 2012-03-16 |
|
|
What does the system user alignment of my PSA EMI Measurement Receiver do?
See contents for details...
FAQ 2012-03-16 |
|
|
Why does my PSA EMI Measurement Receiver tell me that I need to run the user alignment everyday?
There are a couple of reason why this could be hapening. See FAQ contents for details...
FAQ 2012-03-16 |
|
|
What should I do if the system user alignment of my PSA EMI Measurement Receiver fails?
This could be caused by a few different reasons. See contents for details...
FAQ 2012-03-16 |
|
|
What causes the "Final IF Overload" message and how do I correct for the message?
This message occurs when the Final IF is overloaded. This occurs at a mixer level > -8 dBm.
FAQ 2012-02-27 |
|
|
My PSA Series Spectrum Analyzer Shows error +605 Data Corrupt or Stale, NRAM copy of !....
The non-volatile RAM battery has expired. The replacement battery P/N is 1420-0314C. Click here for additional information.
FAQ 2012-02-26 |
|
|
For the E4406A VSA with the latest firmware and the X-Series Analyzers, why are the displayed results for W-CDMA Modulation Accuracy (Rho) different?
Previously with PSA A.06 firmware (E4406A A.07 firmware), mod accuracy only supported 1 slot capture. The current firmware now supports capturing a full frame, and this is the new default setting.
FAQ 2012-02-26 |
|
|
Can I make Noise Figure Measurements below 10 MHz with the ESA-E Series Spectrum Analyzer?
Yes, you can make useful noise figure measurements with the ESA-E Series Spectrum Analyzer down to 200 kHZ, however, no instrument specifications are provided.
FAQ 2012-02-25 |
|
|
手持式频谱分析仪的用户界面是否支持多种语言?
是,每个 HSA 型号的用户界面支持 11 种语言。
FAQ 2012-02-20 |
|
|
能否改变ENA上功能键菜单的字体大小?
能。可通过以下步骤改变功能键菜单字体的大小…
FAQ 2011-10-07 |
|
|
把Agilent Intuilink 4395A连接到4395A, 4396B分析仪上发生异常
错误是由于分析仪的内存RAM磁盘被格式化为LIF引起的。把磁盘重新格式化为DOS驱动器,应用程序将正确运行。
FAQ 2011-06-28 |
|
|
我使用 Agilent 4395A /4396B 组合分析仪的频谱分析仪模式(包括噪声格式)时,把单位设置为 V(伏特),但游标显示为‘nV / Hz’。在此模式下,显示的单位是否有误?
FAQ 2011-06-27 |
|
|
当升级ESA频谱分析仪的最新固件版本时,为什么出现需要选件B72的提示?
固件版本A.08.02及之后的版本需要更大的内存量加载到ESA频谱分析仪。
FAQ 2011-06-24 |
|
|
E4980A/ 4263B/ 4284A/ E4981A 无法对特殊的测试夹具进行开路/短路补偿时,该如何检查测试夹具中的异常情况?
FAQ 2011-06-06 |
|
|
当 4263B 上显示“OVLD”信息时,我应做什么?
当被测阻抗超出量程时,LCD 上就会显示“OVLD”提示。
FAQ 2011-04-21 |
|