E5071C-TDR Method of Implementation (MOI) for High Speed Digital Applications

FAQ 2013-05-21

ENA 6 시리즈는 주파수 스윕할 때 swept sweep 모드를 지원합니까?
아니요..

FAQ 2013-04-25

Replacing the 8508A vector voltmeter with the E5061B network analyzer
Measuring the phase difference of an external CW signals is a typical measurement application of the 8508A vector voltmeter.

FAQ 2013-04-25

[ENA Series] What is the size/type of storage?
The storage type and size are different depending on the HDD revision. See the table.

FAQ 2013-02-04

PDF PDF
Is it possible to programmatically receive the displayed low/high stimulus values when using the “CALC1:MARK:BWID:DATA?” bandwidth query results?
When manually configuring the ENA network analyzer via the Marker – Marker Search – Bandwidth ON, the manual process results in low and high stimulus values as displayed on the analyzer’s front panel.

FAQ 2012-10-31

ENA-L은 DC 레벨을 측정하기 위한 AUX 입력이 있나요?
아니요.

FAQ 2012-10-22

ENA-L에서 871x와 같은 Alt sweep ON/OFF가 가능한가요?
아니요. ENA-L는 항상 Alt sweep ON 입니다.

FAQ 2012-10-22

What is the difference between an "ECal Confidence Check" and a Network Analyzer "System Verification" test?
The ECal Confidence Check is a method to check the accuracy of a calibration performed with either a mechanical cal kit or an ECal module. System verification testing provides network analyzer users a method to verify that the network analyzer is ...

FAQ 2012-10-17

Is it possible on the Agilent ENA series Network Analyzers to over-ride the output source blanking at every sweep end?
Yes, the output source blanking on sweep retrace can be prevented by turning off the Initial Source Port Control function.

FAQ 2012-10-17

Are there any step-by-step measurement examples for ENA 7 series?
Yes. There are some measurement examples in ENA help file...

FAQ 2012-10-17

When should I perform system recovery?
When you encounter the following problems, it's necessary to perform system recovery...

FAQ 2012-09-25

Can I make absolute measurements on the ENA?
Yes.

FAQ 2012-09-25

What mechanical calibration kits and Ecal (electronic calibration) kits does the ENA support?
Please refer to following documents...

FAQ 2012-09-25

What features are different between the ENA 7 series and the ENA 6 series?
Followings are the difference between the ENA 7 series and the ENA 6 series.

FAQ 2012-09-25

What is the cutoff frequency of the rear AUX input ports?
1V range: 10 kHz, 10V range: 2.7 kHz

FAQ 2012-09-25

What are the CPU clock speed and the memory size of the ENA?
Refer to following data. (as of July 2012)

FAQ 2012-09-25

On the ENA 6 series, can I use Supr Avoid ON/OFF function like the 871x?
No. It is always ON.

FAQ 2012-09-25

Is there 85039B calibration definition file for ENA?
Yes. 85039B cal kit definition is avaialble with the firmware revision 9.1 or above. You can also download the file from this page.

FAQ 2012-09-25

Opt.100 VBA utility programs do not run properly. What is wrong?

FAQ 2012-08-09

2 포트 ENA에서 mixed 모드 S-파라미터 측정이 가능 합니까?
...of mixed mode S-parameters on the 2-port ENA

FAQ 2012-07-29

ENA를 오류 계수 어레이에 대한 사용자 액세스를 허용합니까?

FAQ 2012-03-18

On the ENA 6 series, can I run a VBA program which is made on the ENA 7 series?
A VBA project for the ENA 7 series doesn't work on the ENA 6 series as is. However, the program is likely to work with exporting and importing of the modules.

FAQ 2012-03-18

On the ENA 6 series, is the User Defined BEGIN key available like the 871x?
The "Load & Run" of the E5061A/62A and E5061B are equivalent to the User Defined BEGIN key.

FAQ 2012-03-16

ENA에서 "Continuous switching may damage source attenuator"라는 메시지가 표시됩니다. 어떤 추가 조치를 취해야 합니까?
다중 채널 측정 설정에서 다른 전력 범위를 선택할 경우 "Continuous switching may damage source attenuator"라는 메시지가 표시됩니다. 이 경우 과도한 스위칭으로 인해 소스 감쇄기가 손상될 수도 있음을 사용자에게 알리기 위해 오류 메시지가 표시됩니다.

FAQ 2012-03-16

On the ENA 6 series, can I get TTL signal indicating Pass/Fail of the limit test like the 871x?
Yes.

FAQ 2012-03-16