N8212A Performance Vector Upconverter Synthetic Instrument Module, 250 kHz to 20 GHz

Sold By:

Prices for: Italia

* Prices are subject to change without notice. Prices shown are Manufacturer's Suggested Retail Prices (MSRP).

Description

LXI Logo

Agilent's N8212A performance vector upconverter synthetic instrument module is a vector microwave source with greater than 1 GHz I/Q modulation bandwidth to generate a stimulus signal by upconverting a baseband signal. It features superior AM, FM, phase and pulse modulation capabilities (external or internal modulation). The N8212A is based upon the PSG vector signal generator.

Agilent's synthetic instrument family offers the highest-performing RF/MW LAN-based modular instrumentation and the smallest footprint for automated test systems (ATSs); providing the maximum flexibility, and minimizing the cost of an ATS over its lifetime. See a photo of the small modular size mounted in a rack.

Features

  • LXI class-A compliant
  • Similar performance to the E8267D 20 GHz vector signal generator
  • Coherent LO input/output ports allowing use of a common LO signal with multiple upconverters to provide coherent upconversion capability
  • CW, digital step, or analog sweeps
  • Standard high output power and step attenuator
  • Optional narrow pulse capability; enhanced phase noise performance
Specifications 
Frequency range:250 kHz to 20 GHz with .001 Hz resolution
Sweep types:CW, digital step, or analog ramp
Numer of points:2 to 65535, or 2 to 1601 per table
Triggering:Auto, external, single, LAN, or trigger bus
Output power:-20 to +16 dBm typical (for 250 kHz to 20 GHz)
Narrow pulse:(Option UNW) decreases output power by typically 3 dB
Step attenuator range:0 dB and 5 to 115 dB in 10 dB steps
Power level accuracy:1±0.6 dB to ±1.0 dB
Leveling modes:Internal and external detectors
Spectral purity:< -55 dBc typical (2 to 20 GHz)
< -50 dBc typical (20 to 40 GHz)
Digital I/Q modulation bandwidth:1 GHz
FM modulation bandwidth:DC to 1 MHz (typical)
AM modulation bandwidth:DC coupled, 0 to 100 kHz
AC coupled, 10 Hz to 100 kHz, usable to 1 MHz
Phase modulation:10 to 320 radians, normal bandwidth mode
1 to 32 radians, high bandwidth mode
Pulse modulation:2* Refer to following 5 lines
On/off ratio:80 dB
Rise/fall times:6 ns (typical)
Minimum pulse width:0.5 µs (typical)
Pulse repetition frequency:DC to 1 MHz
Pulse modulation level accuracy:±0.5 dB
1 Dependent on frequency and power level range.
2 For frequencies above 3.2 GHz.
 
Ordering Information 
N8212AVector upconverter synthetic instrument module
N8212A-520Frequency range from 250 kHz to 20 GHz
N8212A-UNUStandard pulse modulation
N8212A-UNWNarrow pulse modulation; Requires a U.S. Export License
N8212A-UNXEnhanced phase noise performance; Requires a U.S. Export License
N8212A-UNTAM, FM, PM modulation
N8212A-007Analog ramp sweep interface (does not support 8757)
N8212A-016Wideband, differential, external I/Q inputs; Requires a U.S. Export License