N9330B Handheld Cable and Antenna Tester Firmware

  1. Download firmware upgrade package
  2. Extract and copy N9330B.bin file to the ROOT directory of USB memory stick
  3. Insert the USB stick to the port on the top panel of the instrument
  4. Press {SYS} > {Upgrade} to upgrade the firmware
  5. Cycle the power to N9330B when you see “Upgrade Complete. Effective after Restart.”

    Notice: When upgrading the N9330B firmware from an older version, unrecognized fonts will be displayed instead of this message, resolve this by restarting the N9330B.
  6. Repeat the step 4 and 5.  When the unit is powered on, press {SYS} > {More} > {Status} to check the firmware version, it should be B.01.34.

    Notice: If the display is black when N9330B is powered on, please adjust the “Contrast”, by pressing {Contrast}, and then turn the knob, to see the display information.
Date de mise en circulation Version Description de la version
2011-09-24 B.01.34 Firmware file available for download.

Comment télécharger

  • See Above

Comment installer

  • See Above

Autorisation d'exploitation

  • Ne requiert aucune licence

Document d'appui

Téléchargez

S'installe sur:

Instrument

Préalables

  • See Above

Instruments appuyés

  • N9330B