53181A/31A/32A: Can options be retrofitted on the 53131A/53132A/53181A counters?
This table explains which options may be installed and/or require a service center.
2009-02-25 FAQ
Are the Agilent 53131A, 53132A, and 53181A inputs floating or grounded?
The counter inputs are not floating -- the BNC shield (shell) is at chassis ground (earth ground).
2002-09-09 FAQ
Does the frequency counter come standard with a 10MHz external time base option?
Yes. A standard time base is provided that can be routed to other instruments for synchronization. You may also supply your own external time base as an input to the frequency counter.
2009-03-13 FAQ
How do I configure RS-232 for communications on my 53131/32A or 53181A frequency counter?
None of the counters can communicate over RS-232. Frequency counters must use GPIB to communicate with the PC. RS-232 is used for printing only.
2004-04-14 FAQ
How do I get Instrument Drivers for Agilent Test & Measurement Products?
From the CD shipped with the instrument or from the product technical support web page for the instrument.
2009-05-01 FAQ
How do I rack mount two instruments of different depths?
There are three ways to rack instruments of different depths side-by-side; for example, an Agilent 33220A with an Agilent 34401A.
2007-01-29 FAQ
How do you know when the counter has completed a measurement?
One of three methods: reading the register, reading the Status Byte or generating an interrupt.
2008-03-17 FAQ
How do you prevent the counter from hanging up the GPIB bus when no signal is applied?
Set the GPIB timeout value to a sufficient value to complete the measurement.
2008-03-17 FAQ
What are the hysteresis band values for the Agilent 531xxA-Series Counters?
Some of the counter accuracy formulas (see the manual's specifications section) ask for hysteresis band information.
2005-12-15 FAQ
Where can I find manuals for discontinued and obsolete Hewlett-Packard and Agilent test equipment?
Most Agilent products have PDF manuals that can be found on product pages. Other vendors also supply old manuals.
2009-11-04 FAQ
Why is the fan on the Counter still operating after tuning the power off?
This is to keep the internal Time Base (10 MHz oscillator) stable.
2006-06-20 FAQ