4339B: Why are the 16008B and 16339A equipped with safety cover?
The 16008B and 16339A are equipped with safety cover to avoid electrical shock hazards in compliance with safety regulations.
|
2008-09-11
|
FAQ
|
|
|
|
|
Can I extract the maximum/minimum values from among the measured data in averaging mode?
No, it is impossible.
|
2008-09-11
|
FAQ
|
|
|
|
|
Can I read the offset current data of OPEN correction and reload it to the 4339B over the GPIB?
You can read the offset current data of OPEN correction over GPIB, but cannot reload it in the 4339B.
|
2008-06-10
|
FAQ
|
|
|
|
|
Can I use an external DC source with the 4339B?
Yes. The 4339B can measure the DUT current or resistance whichever the internal DC source or an external DC source is used.
|
2008-06-10
|
FAQ
|
|
|
|
|
Can I use non-Agilent resistivity test fixtures with the 4339B?
Yes, as long as the resistivity test fixture has connectivity to the 4339B, it can be used with the 4339B. It is recommended that the 16117C test leads be used to connect a non-Agilent test fixture.
|
2008-06-10
|
FAQ
|
|
|
|
|
Can the 4339B be used for HIPOT (High-potential) tests?
No, the requirements of HIPOT tests are different from the insulation resistance tests and measurements.
|
2008-09-11
|
FAQ
|
|
|
|
|
How can a normal type BNC connector be connected to the 4339B's high voltage BNC connector?
Normal type BNC connector should not be used because it does not have the withstanding voltage rating enough for the application of 4339B's test voltage output up to 1000 V dc.
|
2008-09-11
|
FAQ
|
|
|
|
|
How can I get the information on the "Interlock connector" circuit of the 4339B?
The Interlock signals and connector pin assignments are described on the 16117C Operation and Service Manual.
|
2008-09-11
|
FAQ
|
|
|
|
|
How do I get Instrument Drivers for Agilent Test & Measurement Products?
From the CD shipped with the instrument or from the product technical support web page for the instrument.
|
2009-05-01
|
FAQ
|
|
|
|
|
How does the Measurement Sequence Mode of the 4339B operate?
This measurement mode automates the measurement sequence of Trigger -> Test Voltage ON -> Wait (Charge Time) -> Measure -> Test Voltage OFF.
|
2008-06-12
|
FAQ
|
|
|
|
|
How often should we perform the verification procedure described in the Performance Tests section of the manual?
Every 12 months.
|
2008-01-11
|
FAQ
|
|
|
|
|
How wide range of resistivity can the 4339B measure when used with the 16008B?
The 4339B can measure surface resistivity up to 7 x 1017 Ω and volume resistivity up to 8 x 1018 Ω-cm, but it depends on the electrodes of the 16008B used.
|
2008-06-11
|
FAQ
|
|
|
|
|
I can't order 16340A CR BOX which is necessary for the performance test of 4339B and 4349B. How can I get 16340A?
The replacement CR box is the combination of 16353A, 16353H and 42038A.
|
2008-07-04
|
FAQ
|
|
|
|
|
Is the 4339B protected against possible short-circuit of measurement terminals when a high voltage is output?
Yes. The 4339B has a current limit feature to protect both DUT and measurement input circuit against possible damages.
|
2008-06-10
|
FAQ
|
|
|
|
|
Is there any safety feature to prevent an operator from getting electrical shocks from the high voltage DC source of the 4339B?
All the test fixtures available for the 4339B have protective features for operator safety.
|
2008-09-11
|
FAQ
|
|
|
|
|
Is there any test fixture for the 4339B to measure SMD components at test voltages above 100 V dc?
The 16339A Component Test Fixture is available.
|
2008-06-10
|
FAQ
|
|
|
|
|
What are the tips on appropriate open correction for measurements using the 4339B with the 16339A?
|
2008-09-11
|
FAQ
|
|
|
|
|
What conditions cause the 4339B to return the measurement status code 5 in response to "FETCh?" command
Refer to Note in Appendix C of the 4339B Operation Manual for the definition of measurement status codes.
|
2008-09-11
|
FAQ
|
|
|
|
|
What does the instrument's rear panel look like?
|
2002-06-27
|
FAQ
|
|
|
|
|
What is input impedance of the 4339B?
Nominal input resistance is 1000 Ω .
|
2008-06-10
|
FAQ
|
|
|
|
|
What is the difference between the 4339A and 4339B?
The 4339B is the enhancement model of the 4339A. Refer to the primary enhancements.
|
2008-09-11
|
FAQ
|
|
|
|
|
What is the signal source output impedance?
|
2002-06-27
|
FAQ
|
|
|
|
|
What is the signal source output resistance of the 4339B?
The output resistance of DC test voltage source is 1 k ohms ±10 %.
|
2008-09-11
|
FAQ
|
|
|
|
|
What should I do when the "OVLD" message appears on my 4339B?
"OVLD" is displayed on the LCD when the measured impedance is out of the measurable range.
|
2008-08-01
|
FAQ
|
|
|
|
|
When does the 4339B display "OVLD"?
|
2008-09-11
|