Are the adapters in the 85033E 30kHz to 9GHz calibration kit phase matched?
Yes.
|
2007-07-10
|
FAQ
|
|
|
|
|
Are there any reference documents for TRL Calibration of network analyzers?
|
2005-05-31
|
FAQ
|
|
|
|
|
Does the Agilent PNA family support programmatic recall of the user preset condition?
Recall of the user preset is either via the SCPI command SYST:UPReset or the MMEM:LOAD *.STA file command, depending on the PNA firmware.
|
2009-03-18
|
FAQ
|
|
|
|
|
ENA & PNA Series Network Analyzer Frequently-Asked Questions (FAQs)
|
2004-11-09
|
FAQ
|
|
|
|
|
For Cable Testing should I use a TDR, a network analyzer, or both?
The cable specifications will determine the test equipment requirements...
|
2009-04-28
|
FAQ
|
|
|
|
|
How can I calculate the equivalent phase delay, as a function of frequency, for a given electrical delay as set on an Agilent Vector Network Analyzer?
Equivalent phase delay, based upon electrical delay is computed as: θ˚ = Ftest * Delay * 360
|
2009-11-10
|
FAQ
|
|
|
|
|
How do I get Instrument Drivers for Agilent Test & Measurement Products?
From the CD shipped with the instrument or from the product technical support web page for the instrument.
|
2009-05-01
|
FAQ
|
|
|
|
|
How is trace math implemented within the Agilent Technologies Vector Network Analyzers?
|
2009-04-13
|
FAQ
|
|
|
|
|
How to do I control an external instrument connected to the GPIB of an Agilent PNA Vector Network Analyzer?
|
2009-03-17
|
FAQ
|
|
|
|
|
I am attempting to program an Agilent PNA, PNA-L, PNA-X Series vector network analyzer but I continue to receive PNA Specific SCPI Error +103. Can you advise?
|
2008-08-28
|
FAQ
|
|
|
|
|
Is it possible to measure the velocity of propagation utilizing an Agilent Vector Network Analyzer (VNA)?
Agilent VNAs do not directly provide the velocity of propagation measurement. As an alternative the steps below outline a method to calculate the velocity of propagation based upon the physical length of the cable under test.
|
2006-06-22
|
FAQ
|
|
|
|
|
Is it possible to perform Mixer Measurements when the Agilent PNA or PNA-L Series Network Analyzer is configured with Option 080 only?
Yes, it is possible to test frequency translating devices (FTDs) on a PNA or PNA-L Series analyzer equipped with option 080 only.
|
2006-09-25
|
FAQ
|
|
|
|
|
Is it possible to set my measurement format to 'Z:Refl' (Z reflection) on my Agilent PNA / PNA-L Series Network Analyzer?
This data may be displayed on the PNA / PNA-L via the new Equation Editor function...
|
2007-05-02
|
FAQ
|
|
|
|
|
Is the Agilent 85070E Material Measurement software compatible with legacy Agilent 85070A/B/C/D High Temperature Dielectric Probes?
Yes, the Agilent 85070E software is backwards compatible with all the legacy high temperature probes Agilent has sold.
|
2006-12-19
|
FAQ
|
|
|
|
|
Is the Agilent Intuilink PNA application supported while the analyzer's frequency offset mode (option 080) is activated?
Agilent Intuilink PNA is not supported when the analyzer is in the frequency offset mode.
|
2009-02-10
|
FAQ
|
|
|
|
|
Is there a process for configuring the Agilent PNA Series Network Analyzer to display Impedance-versus-Time?
Yes. The following outlines the process for configuring the Agilent PNA Series vector network analyzer to indicate impedance as a function of time.
|
2007-10-19
|
FAQ
|
|
|
|
|
My Agilent PNA Family Network Analyzer no longer recognizes the external multiport test set. How do I fix this anomaly?
|
2009-02-12
|
FAQ
|
|
|
|
|
My simulation software is failing to import Touchstone files with extensions *.s1, *.s2, etc. Is there a way to rectify this issue?
|
2009-02-24
|
FAQ
|
|
|
|
|
On my Agilent PNA / PNA-L Series Network Analyzer how do I identify the analyzer's 'Host ID Number'?
On PNA / PNA-L firmware revisions greater than A.06.00 the Host ID is listed under the Help tab.
|
2006-11-01
|
FAQ
|
|
|
|
|
On the Agilent PNA Series Network Analyzers, is it possible to create an ASCII text based data file from a memory trace?
Yes, it is possible to save an Agilent PNA memory trace to an ASCII text file such as a PRN file.
|
2008-10-30
|
FAQ
|
|
|
|
|
Via the Agilent IntuiLink Network Analyzer Application, can I save S2P data files as Logarithmic Magnitude and Phase?
There are no provisions for Log Mag and Phase SxP file save formats.
|
2005-03-30
|
FAQ
|
|
|
|
|
What are my options for running the Agilent 85070E / 85071E software to the PNA Family analyzer?
The Agilent 85070E / 85071E applications support control and communication of the PNA Family analyzers via COM or DCOM only, GPIB is not supported.
|
2006-07-18
|
FAQ
|
|
|
|
|
What are the equations that relate Real and Imaginary data pairs (as read from either a disk save, the GPIB interface, or a display marker) to Log Magnitude, Polar, Phase, and Smith Chart formats?
|
2004-12-13
|
FAQ
|
|
|
|
|
What is the Agilent Part Number for the Agilent PNA Series Network Analyzer Bias Tee Input Fuse?
The replacement part number for the PNA Series Network Analyzer bias tee input fuse is Agilent part number 2110-0046.
|
2006-08-11
|
FAQ
|
|
|
|
|
What is the difference between "sweep speed", "cycle time" and "time per data point" as related to a vector network analyzer?
|
2009-02-24
|
FAQ
|
|