Technical Support:
DSO81204B 80000B Series Infiniium High Performance Oscilloscope: 12 GHz

Product Status: Discontinued | Currently Supported
Product Upgrades: Hardware, Software & Firmware Upgrades
Replacement Product:

DSO91204A Infiniium High Performance Oscilloscope

FAQs

1-20 of 20

Sort:

Results

Title/Description Date Type
Can I install the Infiniium Help system on my PC?
Yes. You may download the Help file for your model of Infiniium and install it on your PC’s hard drive for easy access.
2007-01-10 FAQ
Can I program a waveform captured by my oscilloscope into an Agilent arbitrary waveform generator?
Yes. The free IntuiLink Waveform Editor software will download waveform data from Agilent Infiniium, 6000 series, and 54600 series oscilloscopes and send the data to the an Agilent 33120A, 33210A, 33220A, or 33250A generator.
2008-08-29 FAQ
Does the 1152A active probe work on the DSO80000 Series Infiniium Oscilloscopes?
Yes, the 1152A active probe works with Infiniium oscilloscopes up to and including the new DSO80000B Series oscilloscopes.
2006-11-06 FAQ
For the InfiniiMax probes, what effect does the "Signal being probed" setting have? Why do I have to specify single-ended or differential?
Infiniium scopes and InfiniiMax differential probe amplifiers have a unique method of applying DC offset which considers whether the signal being probed is single-ended or differential...
2006-12-04 FAQ
How do I get my Infiniium to recognize new daylight savings time dates?
For Infiniium oscilloscopes with the Windows® XP operating system, you can download an update from Microsoft ...
2007-02-16 FAQ
How do I get the maximum available memory depth from my Infiniium oscilloscope?
The various Infiniium models offer between 32K and 128M points of memory per channel. The following guide shows you how to get the maximum available memory from your particular scope.
2007-02-22 FAQ
How do I order the Infiniium oscilloscope hard drive replacement kit?
For Infiniium oscilloscopes with the Windows® XP operating system, the part number depends on the oscilloscope model number ...
2008-10-10 FAQ
How do I put an Infiniium oscilloscope into XY mode?
All Infiniium models have a math function called "versus", which plots one channel or function versus another channel or function.
2009-02-06 FAQ
How often should I send in my oscilloscope for calibration?
2006-12-04 FAQ
My InfiniiMax probe fails calibration or does not accurately pick up the cal signal when I use the E2655A test fixture. What could be wrong?
The first step in troubleshooting a failed probe calibration should be to compare the actual signal and the probed signal. If the probed signal does not track the actual signal well, it is most likely due to a poor connection.
2006-06-12 FAQ
My Infiniium oscilloscope with the removable hard drive will no longer boot up, why?
Make sure the removable drive key lock is in the locked position.
2006-10-20 FAQ
My Infiniium scope application is not running properly. Is there a way to reset it?
2007-01-05 FAQ
What is the Agilent part number for the CR2032 3V Lithium battery used on the CPU board in the 5483x, 5485x, MSO8000, and DSO80000 Series Infiniium oscilloscopes?
The Agilent part number is 1420-0356. However, this is a very common battery and may be available at camera shops, electronic suppliers, etc.
2006-11-13 FAQ
What is the amplitude accuracy of my oscilloscope?
When oscilloscope manufacturers specify vertical accuracy, they typically only mention bandwidth and DC accuracy. The bandwidth specification addresses the AC accuracy. Tighter AC accuracy specifications are very uncommon.
2006-11-04 FAQ
What replacement parts and accessories are available for the 113xA-series InfiniiMax probes?
2006-10-18 FAQ
When using TV trigger mode with an Agilent oscilloscope, can I trigger only on even or odd fields?
Yes. All current Agilent scopes with TV triggering are capable of exclusively triggering on odd or even fields.
2006-12-04 FAQ
Where can I find manuals for discontinued and obsolete Hewlett-Packard and Agilent test equipment?
Most Agilent products have PDF manuals that can be found on product pages. Other vendors also supply old manuals.
2009-11-04 FAQ
Why do I see more noise on my oscilloscope when a probe is connected?
2006-12-04 FAQ
Why does the DC Gain Error formula in the DC Gain Accuracy procedure have factors of 0.60 and 0.75 in the DSO80000 Series Infiniiums but not the other Infiniiums?
These factors are used in the DSO80000 Series Infiniiums because the DC gain is measured with input voltages that are less than full scale, but the spec is relative to a full scale input.
2006-10-20 FAQ
Why is the segmented memory acquisition mode useful and which Infiniium oscilloscopes support it?
Segmented memory mode offers more efficient acquisition memory usage for storing waveforms with events of interest that are separated by long periods of time. All currently shipping Infiniiums and many older units support segmented memory.
2006-05-17 FAQ